COMMAND
NAME
DESCRIPTION
NO.
COMMAND SYNTAX
COMMAND
ARGUMENT 1
ARGUMENT 2
ARGUMENT 3 ARGUMENT 4
RESPONSE
K-Factor Read and set K-Factor relative to
N2
Mode:
D – Disabled (K-Factor = 1.0)
I Internal
U – User-Defined
14
K S
(read current settings)
Returns current mode,
index and UD value
KS:Mode,Index,
UDvalue
Example:
KS:U,1,0.91200
D
(Set Mode: Disabled)
KD
I
(Set Mode: Internal Index)
<Index>
[1-22]
KI:Index,GasName
Example: KI:1,Ar
U
(Set Mode: User-Defined Value)
<Value>
[0.00001-999.9]
KU:Value
Example: KU:0.91200
Calibration
Settings
Sets/Reads Calibration related
variables.
Argument 1 = T
Hours since last time unit was
calibrated.
NOTE: must be reset to zero after
calibration.
15 C
P
Flow Power Up Delay
[seconds]
<Value>
[seconds]
CP:<value>
Example: CP:3
No Argument (Returns
Current Power Up Delay)
CP:<value>
Example: CP:3
F
Device Full Scale Range in
liter/min
<Value>
[liter/min]
CF:<value>
Example CF:1000.0
No Argument (Returns
Current Meter FS Range)
CF:<value>
Example: CF:1000.0
L
Device Low Flow Cut Off
in % of full scale
<Value>
(%FS)
CL:<value>
Example: CL:5.0
No Argument
(Returns Current value)
CL:<value>
Example: CL:5.0
T
Read/Reset Calibration/
Maintenance Timer
No Argument
(read timer)
CT:<value>
Example: CT:1024.2
Z
Set Cal. Timer to Zero
CT:Z