Aalborg XFM Digital Manual de usuario Pagina 7

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 56
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 6
2.2.3 Communication Parameters and Connections
The digital interface operates via RS485 (optional RS-232 or Profibus DP avail-
able) and provides access to applicable internal data including: flow, CPU tem-
perature reading, auto zero, totalizer and alarm settings, gas table, conversion
factors and engineering units selection, dynamic response compensation and lin-
earization table adjustment.
Communication Settings for RS-485/RS-232 communication interface:
Baud rate: ...................... 9600 baud
Stop bit: ...................... 1
Data bits: ...................... 8
Parity: ...................... None
Flow Control: ...................... None
RS-485 communication interface connection:
The RS485 converter/adapter must be configured for: multidrop, 2 wire, half
duplex mode. The transmitter circuit must be enabled by TD or RTS (depending
on which is available on the converter/adapter). Settings for the receiver circuit
should follow the selection made for the transmitter circuit in order to eliminate
echo.
RS-485 T(-) or R(-) ...................... pin 8 of the 15 pin "D" connector (TX-)
RS-485 T(+) or R(+) ...................... pin 15 of the 15 pin "D" connector (RX+)
RS-485 GND (if available) ...................... pin 9 of the 15 pin "D" connector (GND)
RS-232 communication interface connection:
Crossover connection has to be established:
RS-232 RX (pin 2 on the DB9 connector) ..... pin 8 of the 15 pin "D" connector (TX)
RS-232 TX (pin 3 on the DB9 connector) ..... pin 15 of the 15 pin "D" connector (RX)
RS-232 GND (pin 5 on the DB9 connector) ..... pin 9 of the 15 pin "D" connector (GND)
4
Vista de pagina 6
1 2 3 4 5 6 7 8 9 10 11 12 ... 55 56

Comentarios a estos manuales

Sin comentarios